Hybrid knee concept combines 3 technologies : hydraulic, pneumatic, electronic.
• Hydraulic shock absorption and a reaction force detecting system (MRS) allow for dynamic control of stance phase. Safety in stance phase is provided by an hydraulic cylinder that adapts the brake to walking conditions.
• The pneumatic unit controls swing phase thanks to a microprocessor that detects changes in tempo and pace and adjusts the speed of extension return, which provides a physiologic walk..
• Hybrid knee meets the needs of active and very active patients by ensuring :
- Stability and control of stance phase during reciprocal walk on slopes and stairs,
- An accurate answer to a wide range of walking paces,
- Free flexion knee in sitting position,
- A one-year battery life.
• Carbon fiber frame.
Adjustments & Technical Points
• Adjustment of damping at the end of extension
• Adjustment of the force and the duration of progressive braking.
• Adjustment of swing phase with an electronic unit.
Knee functioning
• Heel strike :
The MRS system activates the r otative hydraulic unit by increasing the resistance, thus progressively preventing knee flexion (adjustable).
• Stance phase :
Resistance to flexion is adjusted during stance phase to enable the patient to walk stairs and slopes with alternate footsteps.
• End of stance phase :
The MRS system releases resistance to flexion of the rotative hydraulic unit to allow for free flexion and transition into swing phase.
• Swing phase :
A microprocessor measures walking pace and automatically adjusts pneumatic pressure
accordingly, which adapts the prosthesis motions to the walking pace.
Technical features
Item N°. |
1P360 |
1P360-KD |
Weight |
1375 g |
1385 g |
Total Height / Build Height |
309/252 mm |
303/250 mm |
Max. flexion |
140° |
|
Proximal Adapter |
Pyramid |
M36 Thread |
Distal Adapter |
Ø 34 mm tube |
|
Max. Patient Weight |
100 kg |
|
Activity level |
3 - 4 |
|
Battery life |
1 an |